    OT-2 Protocol to transfer volume from several plates to a single plate 

    Ana Mariya Anhel, Lorea Alejaldre, Ángel Goñi-Moreno

    Published: 2024-07-24 DOI: 10.17504/protocols.io.6qpvr4o62gmk/v3
    AI 解读
    浏览 1550

    Abstract

    This protocol is meant to transfer samples from different plates to a single or fewer final plates than the source number, in other words, merge samples of different source plates in a final plate.

    The output of running this script will be the final plate(s) with samples and a reactive (optional), and the corresponding map(s) with the original identities of the samples which will be given by the user in the input file.

    This protocol uses a python script for an Opentrons 2 robot and an excel file containing the required variables to set the number of samples, volumes of transfer, type of plates, etc...

    In our laboratory, this protocol has been used as part of the "High-throughput workflow for the genotypic characterization of transposon library variants" also available in protocols.io

    This protocol is a set of instructions or description of the LAP repository entry LAP-NplateMerging-OT2-2.0.0

    You can find the script and complementary information for this specific version of the protocol in this LAP entry link and GitHub Link to LAP entry documents

    The major changes from previous version are:

    • Change of name variable API Name Rack 15mL Falcon Reactives Name Rack 15mL Falcon Reactives  which is called now  API Name Rack Falcon Reactives
    • The program now accepts falcon tubes and tube racks of 15ml or 50ml. It does not accept mix tube racks.
    • Description of robot and protocol setup in a separate protocols.io entry (Setting and Customizing OT-2 for LAP Entries)
    • The page(s) that corresponds to the map(s) set in the variable 'Name Sheet Map Identifiers' of the page PerPlateVariables need to have the column and row names corresponding to the ones of the labware set in the variable 'API Name Source Plate' of the sheet GeneralVariables

    Before start

    It is important to note the of the source plate in the OT-2 slots, the identity of the samples is provided by excel sheets in the input variable. It should be noted that If 2 samples within a plate have the same name, they will be indistinguishable in the final plate(s)

    Steps

    Files Preparation

    1.

    Preparing Customized Template

    Preparing the template (a .xlsx) with the specific variables for each experiment.

    Attached there is a template of the variable file with several sheets and a PDF file explaining each variable:

    1. GeneralVariables: variables related to the labware to be used
    2. PipetteVariables: variables related to the pipettes to be used
    3. PerPlateVariables: variables related to the specifications of each source plate
    4. Maps (Optional): sheet(s) with the names of the samples in the source plates. These will be reflected in the final plate map --> not included in the template but needs to be included and have the same names as established in the variable Name Sheet Map Identifiers from the PerPlateVariablesSheet Name Sheet Map Identifiers from the PerPlateVariablesSheet

    MergeSamplesInstructionsv200.pdf

    TemplateMergeSamples.xlsx

    Note
    The most updated Excel template can be found in the The most updated Excel template can be found in the LAPrepo Repository Page

    1.1.

    Fill the template with the corresponding values

    1.2.

    Save it with the name VariablesMergeSamples.xlsx

    Note
    The file should be spelt exactly VariablesMergeSamples.xlsx or the Python script won't work correctly

    Setting the robot

    2.

    Prepare the system of the robot to run the protocol

    For this protocol to work we need to transfer the VariablesMergeSamples.xlsx to the directory /data/user_storage of the OT system that we will use to perform the protocol

    As well, if we are using custom labware we need to upload it to the OT App and send it to the directory /data/labware/v2/custom_definitions/custom_beta if the labware is not there yet.

    Finally, we need to make sure the package openpyxl is installed in the robot system

    We can do this entire step by following the protocol Setting and Customizing OT-2 for LAP Entries with the specifications given in the text above

    Setting and Customizing OT-2 for LAP Entries

    Running Protocol

    3.

    Load script in OT-App

    Now that we have transferred the variable files to the robot, we can import the script and run it in the selected robot

    Note
    This whole step has been developed and tested with version 7.0.2 of the OT-AppIndications may vary from version to version of the opentrons App and the version of the script.

    Software

    ValueLabel
    Opentrons AppNAME
    Windows >=10, Mac >=10OS_NAME
    , Ubuntu >=12.04OS_VERSION
    http://opentrons.com/ot-app/REPOSITORY
    OpentronsDEVELOPER
    3.1.

    Load the script in the App

    Protocols -> Import -> Drag Python script

    This version of the protocol was developed when the last version available of LAP-NplateMerging-OT2 was the 2.0.0 which script you can find attached

    ScriptMergePlates_v200.py

    The name of the python file is user's choice, it will work with any name in the app.

    Note
    The last script version can be found at The last script version can be found at http://github.com/BiocomputationLab/LAPrepository/tree/main/LAPEntries. The name of the directory should be LAP-NplateMerging-OT2 followed by the version.. The name of the directory should be LAP-NplateMerging-OT2 followed by the version.As well we can find the latest version of the script at As well we can find the latest version of the script at http://www.laprepo.com/repository/ with the same name as in GitHub with the same name as in GitHub

    Software

    ValueLabel
    LAP RepositoryNAME
    www.laprepo.comREPOSITORY
    http://biocomputationlab.com/DEVELOPER

    Note
    The App with version 7.0.2 analyzes your protocol before setting a robot to run, so the labware will not be shown before assigning the protocol to a specific robot when you import it into the App

    3.2.

    Select Robot to Perform Script

    Click in the protocol -> Start setup -> Choose the OT where the file VariablesMergeSamples.xlsx is -> Proceed To Setup

    After clicking on Proceed to Setup, you should obtain, the positions of the labware in the Labware and in the Liquids tabs, the reagents and their corresponding volume.

    In case the protocol with the set variables cannot run, an error will occur during the run of that simulation. Many errors are contemplated already and have a specific message that hints the user what could have gone wrong.

    Note
    The volume of the initial samples is established to be 90% of the max volume of the well, but it is only a recommendation. Just make sure that there is enough volume to transfer to the final plate(s). The volume of the reagents is exactly what is needed, so it is suggested to pour always more to take in account the error of pipetting

    Note
    It is recommended that you perform a labware position check. You can do it with test plates after loading the script but before cleaning the surface. That way, you reduce the probability of contamination (using the test plates and labware) and pipetting errors (position check).

    4.

    Run Protocol in OT

    4.1.

    Make sure the needed calibrations are done

    Pipettes, tip racks and tip length calibrations need to be done for the items used in this run

    4.2.

    Labware position check is performed (if required)

    4.3.

    Clean the surface of the robot with 70% ethanol to clean and disinfect the surfaces

    Note
    Check the Opentrons page Check the Opentrons page http://support.opentrons.com/s/article/Cleaning-your-OT-2? for more information about cleaning the OT-2 robot with the proper materials for more information about cleaning the OT-2 robot with the proper materials

    4.4.

    Set the labware and reagents as shown in the OT-App

    4.5.

    Start Run

    The procedure that the robot is going to do is mainly divided into 3 parts:

    1. Select which samples, taking into account the variable Type of Sample Selection
    2. (Optional) Distribute reactive(s)
    3. Distribute selected samples

    Citation
    One or more plates , with a reactive mixed with the samples from different sources or only the samples. In addition to an Excel file where every sheet is the map of a final plate located in the folder /data/user_storage /data/user_storage that will give the position in these plates with their identifiers (the name given in the maps of each source plate in the input variable file)

    After-Running

    5.

    Retrieve labware from the OT

    6.

    Import map from robot

    There will be a map with the name set in the variable Name File Final Map in the sheet GeneralVariables followed by the extension .xlsx: [ Name File Final Map ] .xlsx

    This file will be located in the directory /data/user_storage in the robot where the script has been run.

    To retrieve the file, we can and reproduce it by transferring the files from the robot to the computer.

    Take in account that files overwrite, so if you have given the same name to the final map in 2 consecutive runs, you will get only the results of the last run.

    Citation
    The sheets in the final map contain the identity of the samples selected in the places where they have been placed/distributedThese identities are going to be the names given in the map Excel sheets from the input file

    Example

    7.

    We want to merge 16 samples from 5 source plates in a different way, 1 of them selecting samples from back to front, 2 in a random way and 2 from front to back.

    Our final plate will have half the plate already occupied by other samples , and we want to add to all the samples water , so we need to take into account those details

    We will use a computer with a Windows 10 system

    7.1.

    Prepare variable file

    Excel template that we can find filled and saved with the name

    VariablesMergeSamples.xlsx

    VariablesMergeSamples.xlsx

    7.10.

    Retrieve labwares from the OT

    7.11.

    Retrieve the final maps , in this case, the map file will be called example-mapMerge.xlsx (name that is stated in the variable file in the variable Name File Final Map )

    Command line windows with the transfer command of the final plate(s) map(s) file from the OT to our computer
    Command line windows with the transfer command of the final plate(s) map(s) file from the OT to our computer

    example-mapMerge.xlsx

    For more information about how to retrieve files from the robot

    7.2.

    Upload custom labware to app

    We are using a custom labware called vwrblueprintdepth105_96_wellplate_390ul that has been created with the labware creator that opentrons offers (http://labware.opentrons.com/create/)

    vwrblueprintdepth105_96_wellplate_390ul.json

    We upload it to the opentrons app (make sure that is in the robot app) and the robot system as stated in the protocol in step

    vwrblueprintdepth105_96_wellplate_390ul.zip

    (this is a zip file because needed to be compressed to upload to protocols.io but what needs to be transferred to the robot is the folder inside of the zip file)

    7.3.

    Export the variable file to the /data/user_storage folder in the robot the /data/user_storage folder in the robot

    For more information about sending files to the OT-2

    command line window with scp commands to transfer the variables .xlsx from our computer to the OT-2
    command line window with scp commands to transfer the variables .xlsx from our computer to the OT-2
    7.4.

    Import the script that we have downloaded from the step (I named it Example-ScriptMergePlates . py ) to the OT-App

    Example-ScriptMergePlates.py

    Result of importing the Python script into the OT-App
    Result of importing the Python script into the OT-App

    As we can see, we have an error, but that is programmed because the script is meant to work in the robot but not in your computer

    7.5.

    Run the protocol in the robot that we have transferred the Excel file

    Example-ScriptMergePlates.py Example-ScriptMergePlates.py -> Start setup -> Select the robot in which we are going to run the protocol

    If we do not have any errors, the output should look similar to the following pictures

    Labware and liquid set-up layout
    Labware and liquid set-up layout
    Volumes of the needed liquids to perform the protocol
    Volumes of the needed liquids to perform the protocol
    7.6.

    Turn the HEPA filter module

    7.7.

    Clean platform of the robot that we are going to perform the protocol

    7.8.

    Prepare all reagents and labware in the places the App is showing and take into account the notes in step

    7.9.

    Start run

    Citation
    Example of the content of A8 in the labware Final Plate 1 Slot 6
    Example of the content of A8 in the labware Final Plate 1 Slot 6
    Here we will obtain the mix between the volume of the reactive, in this case, water, and the samples selected by the values in the variable file. These positions are seen in the image by the grey wells, and we can see the info of the plate and reactive in the plate on slot 6 in the second picture