• 英国上市公司365

    Total crude protein in plankton: Pierce BCA protein assay (including the enhanced assay for low biomass)

    Ying-Yu Hu, Zoe V. Finkel, Christopher Lord

    Published: 2022-12-12 DOI: 10.17504/protocols.io.5qpvoy5e7g4o/v3
    AI 解读
    浏览 1682

    Abstract

    Here we describe a protocol for extracting total crude protein from phytoplankton and zooplankton, and quantifying by Pierce BCA protein assay. Chlorophyll, phospholipids and sucrose in crude protein could interfere the BCA assay.

    http://assets.thermofisher.com/TFS-Assets/LSG/manuals/MAN0011430_Pierce_BCA_Protein_Asy_UG.pdf

    Steps

    Sample collection

    1.

    Microalgae samples

    1.1.

    Calculate the volume to obtain enough biomass for the assay:

    If using 500 uL extraction buffer, the minimum sampling volume (mL) = 750/(Chl-a_ug/L)

    If using 1000 uL extraction buffer, the minimum sampling volume (mL) = 2X750/(Chl-a_ug/L)

    1.2.

    Filter microalgae in liquid media onto polycarbonate filters, using gentle vacuum pressure (130 mmHg).

    1.3.

    Rinse filter tunnel with filtered artificial seawater (nutrient free) to avoid sample loss.

    1.4.

    Place sample filters in 2 mL Cryogenic Vials.

    1.5.

    Filter blank media (without cells) through polycarbonate filter as blank.

    1.6.

    Flash-freeze tubes with liquid nitrogen and store at -80°C

    2.

    Zooplankton samples

    2.1.

    Grind freeze-dried samples in metal grinding tube (need dry ice)

    Equipment

    ValueLabel
    Metal lysing matrix tubeNAME
    MPBioBRAND
    116992006SKU
    www.mpbio.comLINK

    Equipment

    ValueLabel
    CoolPrep™ adapter for 24 x 2 mL tube holder on FastPrep-24NAME
    MPBioBRAND
    116002528SKU
    www.mpbio.comLINK

    Equipment

    ValueLabel
    FastPrep-24 5GNAME
    Bead-beaterTYPE
    MP BiomedicalsBRAND
    116005500SKU
    2.2.

    Transfer ground sample into Lysing matrix tube, weigh the biomass and log into sampling sheet.

    Equipment

    ValueLabel
    Lysing matrix tube INAME
    MPBioBRAND
    116918100SKU
    www.mpbio.comLINK
    2.3.

    Flash-freeze tubes with liquid nitrogen, store at -80°C until further processing

    3.

    Freeze dry samples before processed.

    Bead tube test for Microalgae samples

    4.

    Bead size and lysing cycles have impact on protein extraction efficiency.

    4.1.

    Bead size

    Note
    Use 2 ml lysing matrix tube for 25 mm filter; 15 ml Teenprep tube for 47 mm filter

    ABCD
    Matrix B0.1Silica spheres116911050-CF
    Matrix Y0.5Yttria-stabilized zirconium oxide beads116960050-CF
    Matrix C1Silica spheres116912050-CF
    Matrix D1.4Zirconium-Silica spheres116913050-CF
    4.2.

    Lysing cycles

    Compare protein yield by using four, six and eight cycles

    4.3.

    Use the optimized bead size and lysing cycles to process protein samples.

    Prepare protein solubilization buffer (PSB)

    5.

    Citation
    Ni G, Zimbalatti G, Murphy CD, Barnett AB, Arsenault CM, Li G, Cockshutt AM, Campbell DA 2017 Arctic Micromonas uses protein pools and non-photochemical quenching to cope with temperature restrictions on Photosystem II protein turnover. Photosynthesis research http://doi.org/10.1007/s11120-016-0310-6

    6.

    In order to obtain compatible results, prepare sufficient PSB so that the same PSB can be used for sample extraction, blank filter extraction and standard solutions

    (1) Extract all samples: Each sample requires 0.25 mL PSB

    (2) Extract all blank filters: Each filter requires 0.25 mL PSB

    (3) Each standard solution (500 ul) requires 0.125 mL PSB

    7.

    For each 10g PSB

    7.1.

    Use anti-statics weighing dish to weigh the following chemicals (one chemical one dish):

    Equipment

    ValueLabel
    Antistatic weighing dishNAME
    FisherbrandBRAND
    08-732-112SKU
    http://www.fishersci.com/us/en/home.htmlLINK

    (1) 0.136g Tris base

    (2) 0.133g Tris HCl

    (3) 0.8g Lithium dodecyl sulphate

    7.2.

    Place a plastic beaker on the top of the scale surface

    7.3.

    Remove the cap of a 15 mL tube and sit it in the beaker

    Equipment

    ValueLabel
    Falcon® Centrifuge TubesNAME
    Polypropylene, Sterile, 15 mLTYPE
    Corning®BRAND
    352096SKU
    7.4.

    Tare the total weight of beaker and tube

    7.5.

    Transfer all chemicals weighed in into the tube, rinse the dish with small amount of MilliQ water to make certain all of the solutes is transferred into the tube

    7.6.

    Use a transfer pipet to add 4g glycerol into the tube

    7.7.

    Add 40µL 0.5Molarity (M) EDTA into the tube

    7.8.

    Top to 10g with MilliQ water

    7.9.

    Vortex until all solutes are completely dissolved.

    Note
    It takes some time to have solutes dissolved, especially LDS. Prepare in advance.

    Prepare Pefabloc solution

    8.

    Note
    Pefabloc is a protease inhibitor, and it loses activity over 24 hours.

    9.

    Add 20.86mL MilliQ into 100mg Pefabloc to obtain a final concentration of 20millimolar (mM).

    10.

    Aliquot into 2.5 mL portions and keep frozen at -20°C

    11.

    The solution can be frozen~thawed multiple times.

    Assay Day 1: Extract protein

    12.

    Prepare protein extraction buffer (PEB):

    Each 1 mL PEB contains

    250 ul PSB

    20 ul 20 mM Pefabloc

    730 ul MilliQ water

    13.

    Prepare ice-bath, keep all samples in the ice-bath

    14.

    Rinse forceps with 70% ethanol and air dry

    Equipment

    ValueLabel
    Filter forcepsNAME
    blunt end, stainless steelTYPE
    MilliporeBRAND
    XX6200006PSKU
    http://www.emdmillipore.com/LINK
    15.

    Label bead tubes and use clean forceps to transfer samples and blank filters into its corresponding bead tube.

    Note
    Bead tube type is selected by

    16.

    Reverse pipet 1mL PEB onto the filter.

    When using 15 mL Teenprep tube, horizontally shake the tube to bury filter into beads before adding PEB, which makes filter easy to be homogenized.

    Note
    Volume of PEB varies due to the actual biomass collected (see guideline)

    17.

    Turn on FastPrep

    Equipment

    ValueLabel
    FastPrep-24 5GNAME
    Bead beaterTYPE
    MP BiomedicalsBRAND
    116005500SKU
    18.

    Check the cap of each tube to make certain cap is tightly screwed. Organize the tubes in order, take notes of the position of each tube, in case the labels get rubbed out during extraction.

    19.

    Run 0h 1m 0s at 6.5 m/s

    20.

    Keep tubes On ice for 0h 1m 0s

    21.

    Check labels. Put tubes back into FastPrep.

    22.

    Run 0h 1m 0s at 6.5 m/s

    23.

    Keep tubes On ice for 0h 1m 0s

    24.

    Check labels. Put tubes back into FastPrep.

    25.

    Run 0h 1m 0s at 6.5 m/s

    26.

    Keep tubes On ice for 0h 1m 0s

    27.

    Check labels. Put tubes back into FastPrep.

    28.

    Run 0h 1m 0s at 6.5 m/s

    29.

    Keep tubes On ice for 0h 1m 0s

    30.

    Note
    More cycles might be required

    31.

    De-foam by centrifuging the extract.

    2 mL Lysing Matrix tubes at 13000rpm,4Room temperature

    15 mL Teenprep tube at 3200x g,4Room temperature

    32.

    Transfer extract

    32.1.

    Microalgae samples in QuickPrep tube (2 mL)

    (1) Transfer all supernatant to a 2 mL microtube

    (2) Centrifuge at 13000rpm,4Room temperature to spin down debris

    (3) Transfer only clear supernatant to a new microtube.

    32.2.

    Microalgae samples in TeenPrep tube (15 mL)

    (1) Use 200 uL tip, go straight to the bottom along the side, try to transfer all extract to a 2 mL microtube.

    (2) Centrifuge at 13000rpm,4Room temperature to spin down debris

    (3) Transfer only clear supernatant to a new microtube.

    32.3.

    Zooplankton samples and Microalgae samples with cloudy extract in which debris is too fine to be centrifuged down

    (1) Use puncher to cut glass fibre filters into about 7 mm disks

    (2) Insert centrifuge filter tube into 2 mL microtube, line the bottom with two glass fibre filter disks by using ethanol rinsed and air-dried tweezers

    (3) Transfer all extract to filter tube

    (4) Centrifuge at 13000rpm,4Room temperature to completely remove debris, and keep the filtrate, discard the filter tube.

    Equipment

    ValueLabel
    Costar® Spin-X® Centrifuge Tube Filters, Corning®BRAND
    33500-692SKU

    Equipment

    ValueLabel
    Microcentrifuge tubeNAME
    CorningBRAND
    29442-590SKU
    33.

    Freeze at -80°C

    Assay Day 2: Prepare Bovine serum albumin (BSA) standard solutions

    34.

    Thaw 20millimolar (mM) pefabloc and transfer 150 ul to a 600 ul microtube.

    Note
    Put the rest of the stock back into the freezer immediately.

    35.

    Thaw extract in the fridge.

    36.

    Organize eight 2 mL microtubes in the tube rack, label the tubes from SD1 to SD8.

    37.

    Reverse pipetting: dispense 125µL PSB into each microtube.

    Note
    When aspiring solution, ensure the pipette to be held vertically. When dispensing, ensure you hold the pipette at an angle (10-45° ). Working to these angles ensures the desired liquid amount is drawn into the tip properly and that all of the liquid is fully dispensed without leaving any residue in the tip.

    http://www.americanlaboratory.com/914-Application-Notes/240482-Ten-Tips-for-Proper-Pipetting/
    http://www.americanlaboratory.com/914-Application-Notes/240482-Ten-Tips-for-Proper-Pipetting/
    38.

    Reverse pipetting: dispense 10µL pefabloc into each microtube

    Note
    Wipe or dab the liquid drop on the outside of the tip, avoid wiping the tip open before dispensing the liquid.

    39.

    Forward pipetting: Add MilliQ into each microtube according to the sheet below:

    ABCDEF
    SD11251036500
    SD21251036050.02
    SD312510353120.048
    SD412510340250.1
    SD512510315500.2
    SD6125102651000.4
    SD7125101652000.8
    SD8125101152501
    40.

    Primary BSA standard

    40.1.

    If BSA (2 mg/mL) is in 50 mL bottle, transfer 1 mL into a microtube.

    40.2.

    If BSA (2 mg/mL) is in ampule, break the ampule with ample opener.

    Equipment

    ValueLabel
    SCIENCEWARE® Break-Safe™ Ampule OpenerNAME
    Bel-Art®BRAND
    89217-378SKU
    41.

    Reverse pipet certain amount of BSA (2 mg/mL) into each tube according to the sheet

    Note
    Wipe or dab the liquid drop on the outside of the tip, avoid wiping the tip open before dispensing the liquid.

    42.

    Vortex each tube.

    43.

    Reverse pipetting: load 4µL of each standard solution onto microdrop plate.

    Equipment

    ValueLabel
    µDrop™ PlatesNAME
    Thermo ScientificBRAND
    N12391SKU
    http://www.lifetechnologies.comLINK
    44.

    Read absorbance of eight standard solutions at 205 nm

    Equipment

    ValueLabel
    Varioskan LUX Multimode Microplate ReaderNAME
    Thermo FisherBRAND
    VL0L00D0SKU
    45.

    Subtract absorbance at 205 nm of blank standard from the 205 nm measurements of all other standard solutions

    46.

    Plot the blank-corrected 205 nm measurement for each standard solution versus its concentration in mg/ml.

    Example of BSA standard curve: Absorbance read at 205 nm versus concentration (mg/mL)
    Example of BSA standard curve: Absorbance read at 205 nm versus concentration (mg/mL)
    47.

    If the standard curve has good Coefficient of Determination, i.e., R2>0.99, the standard solutions are in good quality; otherwise, prepare a new series of standard solutions until the quality of standard solutions meets the requirement.

    48.

    Standard solutions can be kept at Room temperature .

    49.

    Organize eight 2 mL microtubes in the tube rack, label the tubes from SD1 to SD8. Reverse pipet 100µL standard solution into its corresponding tube.

    Assay Day 2: Prepare BCA working reagent (WR)

    50.

    Use the following formula to determine the total volume of WR required. Consider leaving several mL of extra volume:

    (# standards + # samples + # blank filters) X (800µL) = total volume WR required

    51.

    Prepare WR by mixing 50 parts of BCA reagent A with 1 part of BCA Reagent B in a 50 mL falcon tube

    Equipment

    ValueLabel
    Falcon® Centrifuge TubesNAME
    Polypropylene, Sterile, 50 mLTYPE
    Corning®BRAND
    352070SKU

    Assay Day 2: Pierce BCA assay

    52.

    Turn on incubator and preheat to 37°C

    Equipment

    ValueLabel
    SHAKING INCUBATORNAME
    71LTYPE
    Corning® LSE™BRAND
    6753SKU
    53.

    Keep thawed extract On ice

    54.

    Organize 2 mL microtubes in the tube rack, label the tubes for blanks and samples

    55.

    Vortex and then use reverse pipetting: transfer 100µL extract of blanks or samples into the corresponding tubes.

    56.

    Use one tip and reverse pipetting: Add 800µL WR into each tube, make sure that the tip doesn't have contact with the solution, so that samples are not cross-contaminated.

    Note
    Since BCA assay is sensitive to reaction duration, although reagent is aqueous, it is more efficient to use reverse pipetting and quickly dispense reagent into all tubes, therefore the duration difference amongst standards and samples can be minimized.

    57.

    Vortex each tube, shake and incubate at 37°C for 0h 30m 0s

    58.

    Each microplate can hold eight standard solutions and forty samples+blanks, all in duplicate

    Equipment

    ValueLabel
    96-Well MicroplatesNAME
    Polystyrene, Clear,TYPE
    Greiner Bio-OneBRAND
    82050-760SKU
    59.
    Example of organizing samples on the microplate.
    Example of organizing samples on the microplate.
    60.

    Remove samples from the incubator and centrifuge 13300rpm

    61.

    For microplate loading:

    Note
    Reverse pipetting: aspire 200µL sample from the middle of the solutionTip gently touches the side of the well, avoid bending. Dispense 200 uL into the microplateDispose the tipUse a new tip, reverse pipet another 200µL as replicateTip gently touches the side of the well, avoid bending. Dispense 200 uL into the microplate

    62.

    Shake for 5 s at 600 rpm in a continuous and high force modeRead endpoint 562 nm with a measurement time 100 ms

    Equipment

    ValueLabel
    Varioskan LUX Multimode Microplate ReaderNAME
    Thermo FisherBRAND
    VL0L00D0SKU

    Calculate protein content per filter

    63.

    Subtract the average 562 nm absorbance measurement of the blank standard replicates from the 562 nm measurements of all other individual standard .

    64.

    Subtract the average 562 nm absorbance measurement of the blank sample (filter) replicates from the 562 nm measurements of all other individual sample .

    65.

    Prepare a standard curve by plotting the average Blank-corrected 562 nm measurement for each BSA standard versus its concentration in mg/ml.

    66.

    Use the standard curve to determine the protein concentration of each unknown sample by using its blank-corrected 562 absorbance.

    67.

    Calculate the low-limit-of-detection:

    L-LOD_mg/mL=3.3*SD/slope

    where SD is the mean value of standard deviation between each standard replicates.

    L-Abs=L-LOD*slope - intercept

    68.

    If the absorbance of sample is lower than L-Abs, go to Section:

    Assay Day 3: Enhanced Pierce BCA assay for protein 5 to 200 ug/sample

    69.

    Protein_mg/filter = Protein_mg/mL X PEB_mL

    Assay Day 3: Enhanced Pierce BCA assay for protein 5 to 200 ug/sample

    70.
    Response of absorbance at 562 nm to BSA concentration after 30-min incubation at 37 and 60 ºC
    Response of absorbance at 562 nm to BSA concentration after 30-min incubation at 37 and 60 ºC
    71.

    Thaw 20millimolar (mM) pefabloc and transfer 150 ul to a 600 ul mcirotube.

    Note
    Put the rest of the stock back into the freezer immediately.

    72.

    Organize eight 2 mL microtubes in the tube rack, label the tubes from SD1 to SD8.

    73.

    Reverse pipetting: dispense 125µL PSB into each microtube.

    Note
    When aspiring solution, hold the pipet vertically. When dispensing, ensure you hold the pipette at an angle (10-45° ). Working to these angles ensures the desired liquid amount is drawn into the tip properly and that all of the liquid is fully dispensed without leaving any residue in the tip.

    http://www.americanlaboratory.com/914-Application-Notes/240482-Ten-Tips-for-Proper-Pipetting/
    http://www.americanlaboratory.com/914-Application-Notes/240482-Ten-Tips-for-Proper-Pipetting/
    74.

    Reverse pipetting: dispense 10µL pefabloc into each microtube

    Note
    Wipe or dab the liquid drop on the outside of the tip, avoid wiping the tip open before dispensing the liquid.

    75.

    Forward pipetting: Add MilliQ into each microtube according to the sheet below:

    ABCDEF
    SD11251036500
    SD21251036054
    SD312510355108
    SD4125103452016
    SD5125103353024
    SD6125103056048
    SD712510240125100
    SD812510115250200
    76.

    Prepare BSA standard: 0.4mg/mL

    76.1.

    If BSA (2 mg/mL) is in 50 mL bottle, directly reverse pipet 300µL BSA standard into a 2 mL microtube (do not return remaining solution back into the bottle). Forward pipet 600µL+ 600µL Milli-Q into the tube, vortex.

    76.2.

    If BSA (2 mg/mL) is in ampule, break the ampule with ample opener. Reverse pipet 300µL BSA standard into a 2 mL microtube. Forward pipet 600µL+ 600µL Milli-Q into the tube,

    Equipment

    ValueLabel
    SCIENCEWARE® Break-Safe™ Ampule OpenerNAME
    Bel-Art®BRAND
    89217-378SKU
    77.

    Reverse pipet certain amount of BSA (0.4mg/ml) into each tube according to the sheet

    Note
    Wipe or dab the liquid drop on the outside of the tip, avoid wiping the tip open before dispensing the liquid.

    78.

    Vortex each tube.

    79.

    Standard solutions can be kept at Room temperature .

    80.

    Organize eight 2 mL microtubes in the tube rack, label the tubes from SD1 to SD8. Reverse pipet 100µL standard solution into its corresponding tube.

    81.

    Use the following formula to determine the total volume of WR required. Consider leaving several mL of extra volume since Finntip stepper is unable to expel the entire volume from the tip:

    (# standards + # samples + # blank filters) X (800µL) = total volume WR required

    82.

    Prepare WR by mixing 50 parts of BCA reagent A with 1 part of BCA Reagent B in a 50 mL falcon tube

    Equipment

    ValueLabel
    Falcon® Centrifuge TubesNAME
    Polypropylene, Sterile, 50 mLTYPE
    Corning®BRAND
    352070SKU
    83.

    Turn on dry bath and preheat to 60°C

    Equipment

    ValueLabel
    Digital dry bathNAME
    LSETYPE
    CorningBRAND
    6875SBSKU
    http://www.fishersci.com/us/en/home.htmlLINK
    84.

    Keep extracted samples On ice

    85.

    Organize 2 mL microtubes in the tube rack, label the tubes for blanks and samples

    86.

    Forward pipetting: Add 800µL WR into the tubes.

    87.

    Reverse pipetting: transfer 100 µL extract of blanks or samples into the corresponding tubes.

    Note
    Aspire and mix up and down at least three time before transferring the extract

    88.

    Vortex each tube, incubate at 60°C for 0h 30m 0s

    89.

    Each microplate can hold eight standard solutions and forty samples+blanks, all in duplicate

    Equipment

    ValueLabel
    96-Well MicroplatesNAME
    Polystyrene, Clear,TYPE
    Greiner Bio-OneBRAND
    82050-760SKU
    90.
    Example of organizing samples on the microplate.
    Example of organizing samples on the microplate.
    91.

    For microplate loading:

    Note
    Reverse pipetting: aspire 200µL sampleTip gently touches the side of the well, avoid bending. Dispense 200 uL into the microplateReturn remaining sample from the tip back to the tubeDispose the tipVortex the tubeUse a new tip, reverse pipet another 200µL as replicateTip gently touches the side of the well, avoid bending. Dispense 200 uL into the microplate

    92.

    Shake for 5 s at 600 rpm in a continuous and high force modeRead endpoint 562 nm with a measurement time 100 ms

    Equipment

    ValueLabel
    Varioskan LUX Multimode Microplate ReaderNAME
    Thermo FisherBRAND
    VL0L00D0SKU

    推荐阅读

    Nature Protocols
    Protocols IO
    Current Protocols
    扫码咨询